Page 7 of 18                     Report No. 20(04)0336P

                                                 EN 149:2001+A1:2009
             Clause     Requirement + Test                             Result - Remark              Verdict


                          If an exhalation valve is provided it shall be                               N
                          protected against or be resistant to dirt and
                          mechanical damage and may be shrouded or
                          may include any other device

                          Exhalation valve(s) shall continue to operate                                N
                          correctly after a continuous exhalation flow of
                          300 l/min over a period of 30 s.
                          Exhalation valve housing is attached to the                                  N
                          faceblank, and withstand axially a tensile force
                          of 10 N applied for 10 s.
              7.16        Breathing resistance                                                         P
                          The breathing resistances apply to valved and  See 8.9                       P
                          valveless and shall meet the requirements
              7.17        Clogging                                                                     P
              7.17.1      General                                                                      P
                          For single-use devices clogging test is an    Single-use device              P
                          optional test.
                          Devices designed to be resistant to                                          P
                          clogging, shown by a slow increase

                          The specified breathing resistances shall not be                             P
                          exceeded before the required dust load of 833
                                3
                          mg·h/m .
              7.17.2      Breathing resistance                                                         P
              7.17.2.1    Valved particle filtering half masks                                         N
              7.17.2.2    Valveless particle filtering half masks       See 8.9                        P

                          After clogging the inhalation and exhalation  Measured less than 4mbar       P
                          resistances shall not exceed                  Meet the requirement of FFP2
                          -  FFP1: 3 mbar
                          -  FFP2: 4 mbar
                          -  FFP3: 5 mbar
                          at 95 l/min continuous flow.                                                 P

              7.17.3      Penetration of filter materia                                                P
                          All types claimed to meet the clogging                                       P
                          requirement shall also meet the penetration
                          requirements given in 7.9.2 after the treatment.
              7.18        Demountable parts                             No any such part               N
                          All demountable parts (if fitted) shall be readily                           N
                          connected and secured, where possible by
                          hand.



              8           Testing                                                                      P
              8.1         General                                                                      P




                                                                                        TRF No. EN 149_1A