Page 12 of 18                     Report No. 20(04)0336P

                                                 EN 149:2001+A1:2009
             Clause     Requirement + Test                             Result - Remark              Verdict


                         3) moving the head up and down (approx. 15                                    P
                         times), as if inspecting the roof and floor for 2
                         min;

                         4) reciting the alphabet or an agreed text out                                P
                         loud as if communicating with a colleague for 2
                         min;
                          5) walking for 2 min without head movement or                                P
                          talking.
                          g) Record                                                                    P
                          1) enclosure concentration;                                                  P
                          2) the leakage over each exercise period.                                    P

                         h) Turn off the test atmosphere and when the                                  P
                         test agent has cleared from the enclosure
                         remove the subject.
                          After each test, replace the particle filtering half                         P
                          mask by a new sample.
             8.5.2        Method                                                                       P
             8.5.2.1      Principle                                                                    P
                         The subject wearing the particle filtering half                               P
                         mask under test walks on a treadmill over
                         which is an enclosure.

             8.5.2.2      Test equipment                                Meet the requirements          P
                                                                        specified

             8.5.2.3      Expression of results                                                        P
                         The leakage P shall be calculated from         Measured P(%)=6.5%<11%         P
                         measurements made over the last 100 s of       Meet the requirement for
                         each of the exercise periods to avoid carry    FFP2
                         over of results from one exercise to the other.






             8.6          Flammability                                                                 P
                         A total of four particle filtering half masks shall  Two samples as received, and  P
                         be tested: two in the state as received and two  other two samples conditioned
                         after temperature conditioning in accordance
                         with 8.3.2.
                         The single burner test is carried out according to                            P
                         the following procedure.
                         The facepiece is put on a metallic dummy                                      P
                         head which is motorized such that it describes
                         a horizontal circle with a linear speed,
                         measured at the tip of the nose, of (60 ± 5)
                         mm/s.



                                                                                        TRF No. EN 149_1A